Specification
- Quantum efficiency: 82 % (peak QE: @ 560 nm)
- Effective no. of pixels: 2048 (H)×2048 (V)
- Cell size: 6.5 μm×6.5 μm
- Effective area: 13.312 mm×13.312 mm
- Full well capacity: 30 000 electrons (typ.)
- Readout speed: 100 frames/s (Full resolution, standard scan, Camera Link)
- : 40 frames/s (Full resolution, standard scan, USB 3.0, 16 bit)
- : 53 frames/s (Full resolution, standard scan, USB 3.0, 12 bit)
- : 80 frames/s (Full resolution, standard scan, USB 3.0, 8 bit)
- Readout noise: Standard scan (at 100 frames/s, typ.):1.6 electrons rms (1.0 electrons median)
- : Slow scan (at 30 frames/s, typ.): 1.4 electrons rms (0.8 electrons median)
- Cooling method: Peltier cooling
- Cooling temperature: Forced air (Ambient at +20 ℃): -10 ℃
- : Water (+20 ℃): -10 ℃
- : Water (+15 ℃): -30 ℃
- Dark current: 0.06 electrons/pixel/s (Air Cooled to -10° C) (typ.)
- : 0.06 electrons/pixel/s (Water Cooled to -10° C) (typ.)
- : 0.006 electrons/pixel/s (Water Cooled to -30° C) (typ.)
- Dynamic range: 37 000:1 (typ.)
- Interface: Camera Link / USB 3.0
- A/D converter: 16 bit / 12 bit / 8 bit
- Readout modes: Normal Area, Lightsheet, W-VIEW Mode, Dual Lightsheet
- Binning: 2x2, 4x4
- Master Pulse Generator (Pulse Modes): Internal Sync, Start Trigger, Burst
- Master Pulse Generator (Pulse Interval in 1 μs increments): 10 μs to 10 s
- Dark Signal Non-Uniformity (DSNU): 0.3 electrons rms
- Photo Response Non-Uniformity (PRNU) at half level of full light range (15,000 electrons): 0.06 % rms
- Photo Response Non-Uniformity (PRNU) at low light level (700 electrons): 0.3 % rms
- Linearity error, full light range (EMVA 1288 standard): 0.5 %
- Linearity error, low light range (< 500 electrons signal): 0.2 % / Less than approx. 1 electron absolute error